Hydrographic Team Activity Report (Healy 04-02)

15 May to 23 June 2004

Nome, Alaska to Nome, Alaska

Dr. James Swift, (on board PI), Dr. Louis Codispoti (PI)

On board team: Kristin Sanborn, Jennifer Sheldon, Dan Schuller, Doug Masten,

Robert Palomares

Other team members: Dave Huntley (ADCP), Dean Stockwell (Chlorophyll)

98 CTD casts on 35 stations were attempted. One of these was aborted, with no CTD data and no water samples, six additional casts were aborted, the CTD data were reported, but there were no water samples.  These casts were:

                                                Station              Cast

                                                003                  01        CTD data reported, 12 bottles tripped.

006                  02        CTD data reported, 12 bottles tripped.

016                  01        CTD data reported, 4 bottles tripped.

016                  03        CTD data not reported, no bottles

027                  01       CTD data reported, no bottles.

031                  03        CTD data reported, aborted mid down-cast

Instrumentation

CTD casts were performed with a rosette system consisting of a 12-place rosette frame with 30 liter bottles and a 12-place SBE-32 Carousel pylon.  Underwater electronic components consisted of:

·        Sea-Bird Electronics, Inc. (SBE) 911plus CTD,

·        WETLabs C-Star transmissometer with a 25cm path length and 660nm wavelength,

·        Biospherical Instruments, Inc. Photosynthetically Active Radiation (PAR) sensor,

·        Chelsea MkIII Aquatracka fluorometer, and

·        Simrad, 5 volt - 500 meters altimeter.

Additionally, a Dr. Haardt fluorometer designed to detect colored organic matter (CDOM) and a Secchi disk were mounted on the CTD package. The CTD, transmissometer, and the two fluorometers were mounted horizontally along the bottom of the rosette frame. The PAR sensor was located at the top of the rosette. The surface PAR sensor was located on the aft, starboard railing of the helicopter shack.  All sensors except the Secchi disk were interfaced with the CTD system. This instrument package provided pressure, dual temperature and dual conductivity channels as well as light transmissivity and fluorometric signals at a sample rate of 24 scans per second.

The bottles on the rosette were General Oceanic 30 liter bottles. The bottles were equipped with internal nylon coated springs and silicone o-rings which are used to minimize toxicity to the sample. Bottle numbering is 1 to 12 with 1 tripped first usually at the deepest sampling level and 12 tripped last at the shallowest sampling level. The rosette system was suspended from a standard UNOLS 3 conductor 0.322” electromechanical cable.

The CTD used was serial number 09P24152-0638 and the sensor’s model and serial numbers are listed in Table 1.

 

 

TABLE 1. Instrument/Sensor Serial Numbers

 

Primary

Temperature

Primary

Conductivity

Secondary

Temperature

Secondary

Conductivity

Pressure

Transmissometer

SBE 3plus

SBE 4C

SBE 3plus

SBE 4C

401K-105

C-Star

03-2796

04-2545

03-2824

04-2568

83009

CST-390DR

 

Oxygen

Fluorometer

PAR

Surface PAR

Altimeter

SBE 43

Aqua 3

QSP-2300

QSR-240

807

0459

088233

4643

6367

9711090

 

Equipment Positions

TABLE 2. Instrument mounting heights in reference to the bottom of the rosette frame.

 

Sensor

Height above base of rosette

 

Sensor

Height above base of rosette

Altimeter

2 cm

 

Pressure

19cm

Transmissometer

 8 cm

 

T (pri)

10 cm

 

Fluorometer (Chelsea)

10 cm

 

 

Fluorometer (Haardt)

8 cm

 

Par

215cm Sta. < 2000 m

 

 

 

 

 

 

The distance of the mid-points of the 30 L Niskin bottles from the bottom-mounted sensors was ~1.19m. The 30 Liter Niskin bottles are ~1.0 m long. The secchi disk was mounted 2.2m above the bottom of the rosette frame.

Problems and/or Procedural changes

Bottle 7 was replaced after station 010.  At times the nylon coating on the springs broke down and some rust was apparent. To minimize the occurrence of rust, the springs were inspected before the cruise and, as feasible during the cruise. During the mid-cruise servicing of the CTD/rosette system that occurred following station 021, all springs were inspected and 6 were replaced. HLY0402 rosette operations were continually beset by problems with bottle leaks caused by Niskin bottle end o-rings falling out of position.  Typically, each cast had one such occurrence.  Although some Niskin bottles were more prone than others to have an o-ring problem, in general the problem shifted from bottle to bottle between casts.  Some of the problems were gross, i.e. the o-ring would be visible out the side of the end cap, but others were more subtle. Every time an o-ring problem was suspected, the o-ring was carefully inspected, and replaced if necessary.  Also, at several points during the cruise all o-rings were inspected.  The contents of various packages of spare o-rings were measured to locate 'large' or 'small' o-rings (within the manufacturer's tolerance), and a remedial 'large' set was installed.  Another time Coast Guard personnel replaced all the o-rings from their own supply.  Yet all these remedial attempts were to no particular avail.  The problem bears further thought toward a satisfactory solution.

CTD Data

CTD Laboratory Calibration Procedures

Pre-cruise laboratory calibrations of CTD pressure, temperature and conductivity sensors were used to generate coefficients for the calculation of these parameters from their respective sensor frequencies. The temperature and conductivity calibrations were performed at Sea-Bird Electronics, Inc. in Bellevue, Washington.  Calibration of the pressure sensor was performed by Scripps Institution of Oceanography, Shipboard Technical Support/Oceanographic Data Facility (SIO/STS/ODF) personnel. The Sea-Bird laboratory temperature calibrations were referenced to the International Temperature Scale of 1990 (ITS-90).

CTD Data Acquisition

The CTD 911plus was operated generally as suggested in the Sea-Bird CTD Operating and Repair Manual, which contains a description of the system, its operation and functions (Sea-Bird Electronics, Inc., 2002). One difference from Sea-Bird’s operation is that data acquisition was started on deck. This procedure allows a check of the pressure offset and an unblocked reading of the transmissometer. The Seasoft acquisition program as described in the CTD Data Acquisition Software Manual (Sea-Bird Electronics, Inc., 2001) provided a real-time graphical display of selected parameters adequate to monitor CTD performance and information for the selection of bottle-tripping depths. Raw data from the CTD were archived on the PC’s hard disk at the full 24 Hz sampling rate.

A CTD Station Sheet form was filled in for each deployment, providing a record of times, positions, bottom depth, bottle sampling depths, and every attempt to trip a bottle, as well as any pertinent comments. When the equipment and personnel were ready, data acquisition was started.  The CTD operator pressed a control key (flag), which appends a summary line into the files created for “inventory” files.  This file contains a summary of the time, ship’s position, and current scan number each time the control key is pressed.  They are used as a reference to mark important events during the cast, such as on deck pressure, when the lowering was initiated, when the package was at the bottom, when bottles were tripped and the on-deck pressure with ending position.  After the initial flag, the rosette/CTD system was lowered into the water and held at 5 meters wire out for 3-5 minutes to permit activation of the CTD pumps and equilibration of the sensors.  Then, the operator had the CTD raised to the surface,  again created a flag, and simultaneously directed the winch operator to begin lowering.  The operator created a flag at the deepest point of the cast. Bottom depths were calculated by combining the distance above bottom, reported by the altimeter, and the maximum depth of the CTD package when bottom altimeter readings were available.  If there was no altimeter reading, then the bottom depth is reported from the ship’s Bathy 2000 or Knudsen model 320B/R depth recorder.  These data, corrected for the draft of the transducer, were logged in uncorrected meters (assuming a sound velocity of 1500 m/sec). If the altimeter and depth recorder data were unavailable, the final resort was to use depth data from the SeaBeam system (corrected sound velocities).

 

 The wire out corresponding to each bottle trip was written on the station log and the trips were electronically flagged in the data file.  The performance of all sensors was monitored during the cast.  After the rosette recovery, the operator created a final flag denoting the end of the cast. Any faulty equipment or exceptionally noisy data were noted on the log sheet.

Problems and Procedural changes

Prior to station 007, position information was not being appended to every scan. The wrong configuration file was later inadvertently chosen and the absolute positions were not appended to the data for Stations 020 casts 3-7, 021 cast 01, 023 casts 1-1, 024 casts 2-3 and 025 cast 1.

CTD Data Processing

Pressure

CTD values determined on deck before and after each cast were compared to determine a pressure offset correction. The comparison suggested that no pressure offset was necessary.

Temperature

The temperature sensors were calibrated in November of 2003.  The dual temperature sensors were monitored during the expedition and exhibited good agreement.  It appears that no additional corrections need to be applied. A post-cruise calibration will be performed.

Conductivity

Corrected CTD pressure and temperature values were used with bottle salinities to back-calculate bottle conductivities. Comparison of these bottle values with the CTD primary conductivity values indicated no additional offset needed to be applied to the data.

Transmissometer

A WETLabs calibrated transmissometer was utilized throughout the cruise.  An on deck calibration check was performed and even though there was little degradation from the last calibration, the new coefficients were applied to the data set.

Oxygen, Fluorometer, and PAR

The CTD oxygen data are only intended for qualitative use.  Similarly, the fluorometric and PAR data are not calibrated.

Data Processing

Sea-Bird Seasoft CTD processing software was employed. The processing programs are outlined below.  A more complete description may be found in the Sea-Bird Software Manual which is available from the Sea-Bird website (www.seabird.com).

The sequence of programs that were run in processing CTD data from this cruise are as follows:

·        DATCNV - Converts data from raw frequencies and voltages to corrected engineering units

·        WILDEDIT - Eliminates large spikes

·        CELLTM - Applies conductivity cell thermal mass correction

·        FILTER A low pass filter to smooth pressure for LOOPEDIT

·        LOOPEDIT - Marks scans where velocity is less than selected value to avoid pressure reversals from ship roll, or during bottle flushing.

·        DERIVE - Computes calculated parameters

·        BINAVG - Average data into desired pressure bins

The quality control steps included:

·        Sensor verification After the CTD was set up and sensor serial numbers and sensor location was entered into the computer, another check was made to verify that there were no tabulation errors.

·        Seasoft Configuration File was reviewed to verify that individual sensors were represented correctly, with the correct coefficients.

·        Temperature was verified by comparing primary and secondary sensor data.

·        Conductivity was checked by comparison of the two sensors with each other and with bottle salinity samples.

·        Position Check A chart of the ship’s track was produced and reviewed for any serious problems.  The positions were acquired from the ship’s Trimble P-code navigation system.

·        Visual Check Plots of each usable cast were produced and reviewed for any noise and spikes that may have been missed by the processing programs.

·        The density profile was checked for inversions that might have been produced by sensor noise or response mismatches.

 

 

CTD Data Footnoting

WHP water bottle quality flags were assigned as defined in the WOCE Operations Manual [Joyce]. These flags and interpretation are tabulated in the CTD and Bottle Data Distribution, Quality Flags section of this document.

Data Comments

Fine structure including minor density inversions that may appear in the upper ~ 10 m of the profiles is most likely caused by ship discharges/turbulence. To minimize the ship effect, engine cooling water discharges were restricted to the port side of the Healy.  A “yo yo” procedure was adopted to induce bottle flushing whenever waves and ship motion were weak.  This procedure was employed for all bottle trips under quiescent conditions except for productivity casts, and for some thin low salinity surface lenses.  In the latter cases, the CTD was raised slowly so as not to disturb the thin low-salinity surface layers with the CTD wake, and the soak time was relied on to flush the bottle.  Even though this procedure may not have adequately flushed the surface bottle, it was sufficient to reveal some large salinity differences in the ~ 1m depth interval separating the CTD sensor from the bottle mid-point.  These situations occurred in melting ice under low winds and waves, and it is suspected that the water may have been stratified even within the surface Niskin bottle. Regardless of the procedure employed, the CTD operators were instructed to wait for at least 1 minute (typically > 1.5 minutes) before tripping the bottle.

All salinity, nutrient and dissolved oxygen data collected by the “service” team have gone through several stages of editing and are not likely to change significantly.  The chlorophyll observations reported are, however, preliminary and may undergo significant post-cruise editing.

Bottle Data

There were five generic types of casts performed with differing sampling protocols.  Generally speaking, the samplings during these casts were as follows, but there is some cast-to-cast variation.

·        Hydrographic

o       Oxygen,

o       Total CO2,

o       Total Alkalinity,

o       Nutrients

o       Chlorophyll/Phaeophytin

o       Phytoplankton

o       Salinity

o       O18/O16

o       Benthic

o       Dissolved Organic Matter/Particulate Organic Matter

o       Thorium-234

·        Productivity/Zooplankton

o       Oxygen

o        Oxygen Respiration

o       Productivity

o       Nutrients

o       Chlorophyll

o       HPLC

o       Bacteria

o       Micro Zooplankton

o       Particulate Organic Matter

o       Dissolved Organic Matter/Lignin

o       Bio-Optics

o       Taxonomy

o       C13/N15

·        Bio-Markers

o       Nutrients

o       Particulate Organic Matter

o       Dissolved Organic Matter/Lignin

·        Radium

o       Nutrients

o       Radium

·        Zooplankton

o       Nutrients

o       Micro Zooplankton

o       C13/N15

 

The correspondence between individual sample containers and the rosette bottle from which the sample was drawn was recorded on the sample log for the cast. This log also included any comments or anomalous conditions noted about the rosette and bottles.

 

Normal sampling practice included opening the drain valve before the air vent on the bottle, to check for air leaks. This observation together with other diagnostic comments (e.g., "lanyard caught in lid", "valve left open") that might later prove useful in determining sample integrity was routinely noted on the sample log. 

 

Bottle Data Processing

After the samples were drawn and analyzed, the next stage of processing involved merging the different data streams into a common file. The rosette cast and bottle numbers were the primary identification for all ODF-analyzed samples taken from the bottle, and were used to merge the analytical results with the CTD data associated with that bottle.

Diagnostic comments from the sample log, and notes from analysts and/or bottle data processors were entered into a computer file associated with each station (the "quality" file) as part of the quality control procedure. Sample data from bottles suspected of leaking were checked to see if the properties were consistent with the profile for the cast, with adjacent stations, and, where applicable, with the CTD data. Direct inspection of the tabular data, property-property plots and vertical sections were all employed to check the data. Revisions were made whenever there was an objective reason to delete, annotate or re-calculate a datum. WHP water sample codes were selected to indicate the reliability of the individual parameters affected by the comments. WHP bottle codes were assigned where evidence showed the entire bottle was affected, as in the case of a leak, or a bottle trip at other than the intended depth.

Specific data processing and techniques and additional quality control are included with the parameter write-up.

Pressure and Temperatures

All pressures and temperatures for the bottle data tabulation were obtained by averaging CTD data for a brief interval at the time the bottle was closed and then applying the appropriate calibration data.

 

The temperatures are reported using the International Temperature Scale of 1990.

Salinity

384 salinity samples were analyzed in 14 analyses runs.

Sampling and Data Processing

Salinity samples were drawn into 200 ml high alumina borosilicate bottles, which were rinsed three times with sample prior to filling. The bottles were sealed with custom-made plastic insert thimbles and Nalgene screw caps. This container provides very low container dissolution and sample evaporation. 

Equipment and Techniques

A Guildline Autosal 8400B #65-715, standardized with IAPSO Standard Seawater (SSW) batch P-144, was used to measure the salinities. Prior to the analyses, the samples were stored to permit equilibration to laboratory temperature, usually 8-20 hours.  The salinometer was outfitted with an Ocean Scientific International interface for computer-aided measurement. The salinometer was standardized with a fresh vial of standard seawater (SSW) at the beginning of each analysis run.  Instrument drift was determined by running a SSW vial after the last sample was run through the autosal. The salinometer cell was flushed until two successive readings met software criteria for consistency; these were then averaged for a final result. The estimated accuracy of bottle salinities run at sea is usually better than 0.002 PSU relative to the particular standard seawater batch used.

Laboratory Temperature

The temperature stability in the salinometer laboratory was good; variation was no more than 1ºC during a run of samples.  The laboratory temperature was generally 2-3ºC lower than the Autosal bath temperature.

Oxygen

463 samples were analyzed for oxygen.

Sampling and Data Processing

Samples were collected for dissolved oxygen analyses as the first sample after the rosette was brought on board. Using a Tygon drawing tube, nominal 125ml volume-calibrated iodine flasks were rinsed three times, then filled and allowed to overflow for approximately 3 flask volumes. The sample draw temperature was measured with a small platinum resistance thermometer embedded in the drawing tube. Reagents were added to fix the oxygen before stoppering. The flasks were shaken twice to assure thorough dispersion of the precipitate, once immediately after drawing, and then again after about 20 minutes.  The samples were usually analyzed within a few hours of collection.

Thiosulfate normalities were calculated from each standardization and corrected to 20ºC. Periodically, the 20ºC normalities and the blanks were plotted versus time and were reviewed for possible problems. New thiosulfate normalities were recalculated as a linear function of time, if warranted. The oxygen data were recalculated using the smoothed normality and an averaged reagent blank. Oxygens were converted from milliliters per liter to micromoles per kilogram using the sampling temperature.

 

 

Equipment and Techniques

Dissolved oxygen analyses were performed with an ODF-designed automated oxygen titrator using photometric end-point detection based on the absorption of 365nm wavelength ultra-violet light. The titration of the samples and the data logging were controlled by PC software. Thiosulfate was dispensed by a Dosimat 665 buret driver fitted with a 1.0 ml buret. The ODF method used a whole-bottle modified-Winkler titration following the technique of Carpenter (1965) with modifications by Culberson (1991), but with higher concentrations of potassium iodate standard (approximately 0.012N) and thiosulfate solution (55 g/l). Standard KIO3 solutions prepared ashore were run at the beginning of each run. Reagent and distilled water blanks were determined, to account for presence of oxidizing or reducing materials.

 

Volumetric Calibration

 

Oxygen flask volumes were determined gravimetrically with degassed deionized water to determine flask volumes at ODF’s chemistry laboratory. This was done once before using flasks for the first time and periodically thereafter when a suspect bottle volume was detected. The volumetric flasks used in preparing standards were volume-calibrated by the same method, as was the 10 ml Dosimat buret used to dispense standard iodate solution.

Standards

 

Potassium iodate was obtained from Johnson Matthey Chemical Co. and was reported by the supplier to be >99.4% pure.

Nutrients

1229 samples were analyzed for nutrients in 63 analyses runs.

Sampling and Data Processing

Nutrient samples were drawn into 45 ml polypropylene, screw-capped “oak-ridge type” centrifuge tubes. The tubes were rinsed with 10% HCl and then with sample three times before filling. Standardizations were performed at the beginning and end of each group of analyses (typically 6-24 samples) with an intermediate concentration mixed nutrient standard, which was prepared prior to each run from a secondary standard in a low-nutrient seawater matrix. The secondary standards were prepared aboard ship by dilution from primary standard solutions.  Dry standards were pre-weighed at the laboratory at ODF, and transported to the vessel for dilution to the primary standard. Sets of 6-7 different standard concentrations covering the range of sample concentrations were analyzed periodically to determine the deviation from linearity, if any, as a function of concentration for each nutrient analysis.  A correction for non-linearity was applied to the final nutrient concentrations when necessary.

After each group of samples was analyzed, the raw data file was processed to produce another file of response factors, baseline values, and absorbances.  These values were then checked for accuracy against values taken from strip chart recordings. A stable deep seawater check sample was run occasionally as a substandard check.

 

Nutrients, when reported in micromoles per kilogram, were converted from micromoles per liter by dividing by sample density calculated at 1 atm pressure (0 db), in situ salinity, and the sample temperature measured at the time of analysis.

Equipment and Techniques

 

Nutrient analyses (nitrate+nitrite, nitrite, phosphate, silicate, ammonium, and urea) were performed on an ODF-modified 6-channel Technicon AutoAnalyzer II, generally within a few hours after sample collection.  The samples were kept in the dark by covering with tin foil or refrigerated at 4°C, if necessary, but brought to within 5°C of lab temperature before analysis.  The analog outputs from each of the six channels were digitized and logged automatically by computer (PC) at 2-second intervals.

 

A modification of the Armstrong et al. (Armstrong 1967) procedure was used for the analysis of nitrate and nitrite. For the nitrate plus nitrite analysis, the seawater sample was passed through a cadmium reduction column where nitrate was quantitatively reduced to nitrite. The stream was then passed through a 15mm flowcell and the absorbance measured at 540nm.  The same technique was employed for nitrite analysis, except the cadmium column was bypassed, and a 50mm flowcell was used for measurement.  Periodic checks of the column efficiency were made by running alternate equal concentrations of NO2 and NO3 through the NO3 channel to ensure that column efficiencies were high (> 95%). Nitrite concentrations were subtracted from the nitrate+nitrite values to obtain nitrate concentrations.

Phosphate was analyzed using a modification of the Bernhardt and Wilhelms [Bernhardt 1967.] technique. The reaction product was heated to ~55ºC to enhance color development, then passed through a 50mm flowcell and the absorbance measured at 820m.

 

Silicate was analyzed using the technique of Armstrong et al., (Armstrong, 1967). The sample was passed through a 15mm flowcell and the absorbance measured at 660nm.

 

Ammonium was determined by the Berthelot reaction (Patton and Crouch 1977) in which sodium hypochlorite and phenol react with ammonium ion to produce indophenol blue, a blue compound.   The solution was heated to 55°C and passed through a 50mm flowcell at 640nm.

 

 

Urea was analyzed via a modification of the method of Rahmatullah and Boyde (1980), which is based on the classic diacetyl monoxime method.  A solution of diacetyl monoxime, thiosemicarbizide and acetone is followed by the addition of ferric chloride, which acts as a catalyst.  The resultant solution was heated to 90°C and passed through a 50mm flowcell. The absorbance was measured at 520nm.

 

Nutrient Standards

Na2SiF6, the silicate primary standard, was obtained from Johnson Matthey Company and Fisher Scientific and was reported by the suppliers to be >98% pure. Primary standards for nitrate (KNO3), nitrite (NaNO2), and phosphate (KH2PO4) were obtained from Johnson Matthey Chemical Company, and the supplier reported purities of 99.999%, 97%, and 99.999%, respectively. Ammonia, (NH4(SO4)2), and Urea primary standards were obtained from Fisher Scientific and reported to be >99% pure.

Bottle Data Footnoting

WHP water bottle quality flags were assigned as defined in the WOCE Operations Manual [Joyce]. These flags and interpretation are tabulated in the Data Distribution, Bottle Data, Quality Flags section of this document.

Data Distribution

The CTD and bottle data can be obtained through the JOSS website, www.JOSS.ucar.edu/sbi. The data are reported using the WHP-Exchange (WOCE Hydrographic Program) format and the quality coding follows those outlined by the WOCE program (Joyce, 1994). In addition, the format can be obtained through the WOCE Hydrographic Program website, www.WHPO.ucsd.edu. The descriptions in this document have been edited from the reference to annotate the format specific to this data distribution. ASCII files for each station were created with comments recorded on the CTD Station Logs during data acquisition. These ASCII files include data processing comments noting any problems, the resolution, and footnoting that may have occurred.  A separate ASCII file was also created with the comments from the Sample Log Sheets that include problems with the Niskin bottles that could compromise the samples. Comments arising from inspection and checking of the data are also included in the ASCII file. These comment files are also in the JOSS database.

General rules for WHP-exchange:

  1. Each line must end with a carriage return or end-of-line.
  2. With the exception of the file type line, lines starting with a "#" character, or including and following a line which reads "END_DATA", each line in the file must have exactly the same number of commas as do all other lines in that file.
  3. The name of a quality flag always begins with the name of the parameter with which it is associated, followed by an underscore character, followed by "FLAG", followed by an underscore, and then followed by an alphanumeric character, W.

4.      The "missing value" for a data value is always defined as -999, but written in the decimal place format of the parameter in question. For example, a missing salinity would be written -999.0000 or a missing phosphate -999.00.

5.      The first four characters of the EXPOCODE are the U.S. National Oceanographic Data Center (NODC) country-ship code, then followed by up to an 8 characters expedition name of cruise number, i.e. 31H1HLY0402.

CTD Data

Description of 31H1HLY0402_ct1.csv file layout.

1st line

File type, here CTD, followed by a comma and a DATE_TIME stamp

 

YYYYMMDDdivINSwho

 

YYYY   4 digit year

MM     2 digit month

DD     2 digit day

div    division of Institution

INS    Institution name

who    initials of responsible person

 

# lines

A file may include 0-N optional lines at the start of a data file, each beginning with a "#" character and each ending with carriage return or end-of-line.  Information relevant to file change/update history may be included here, for example.

2nd line

NUMBER_HEADERS = n (n = 10 in this table and the example_ct1.csv file.)

3rd line

EXPOCODE = [expocode] The expedition code, assigned by the user.

4th line

SECT_ID = [section] The SBI station specification. Optional.

5th line

STNNBR = [station] The originator's station number

6th line

CASTNO = [cast] The originator's cast number

7th line

DATE = [date] Cast date in YYYYMMDD integer format.

8th line

TIME = [time] Cast time that CTD was at the deepest sampling point.

9th line

LATITUDE = [latitude] Latitude as SDD.dddd where "S" is sign (blank or missing is positive), DD are degrees, and dddd are decimal degrees. Sign is positive in northern hemisphere, negative in southern hemisphere

10th line

LONGITUDE = [longitude] Longitude as SDDD.dddd where "S" is sign (blank or missing is positive), DDD are degrees, and dddd are decimal degrees. Sign is positive for "east" longitude, negative for "west" longitude

11th line

DEPTH = [bottom] Reported depth to bottom. Preferred units are "meters" and should be specified in Line 2. In general, corrected depths are preferred to uncorrected depths. Documentation accompanying data includes notes on methodology of correction. Optional.

next line

Parameter headings.

next line

Units.

data lines

A single _ct1.csv CTD data file will normally contain data lines for one CTD cast.

END_DATA

The line after the last data line must read END_DATA, and be followed by a carriage return or end of line.

other lines

Users may include any information they wish in 0-N optional lines at the end of a data file, after the END_DATA line.

Parameter names, units, format, and comments

Parameter

Units

Format

Comments

CTDPRS

DB

F9.1

CTD pressure, decibars

CTDPRS_FLAG_W

 

I1

CTDPRS quality flag

CTDTMP

ITS-90

F9.3

CTD temperature, degrees C (ITS-90)

CTDTMP_FLAG_W

 

I1

CTDTMP quality flag

CTDSAL

 

F9.3

CTD salinity

CTDSAL_FLAG_W

 

I1

CTDSAL quality flag

CTDOXY

UMOL/KG

F9.1

CTD oxygen, micromoles/kilogram

CTDOXY_FLAG_W

 

I1

CTDOXY quality flag

XMISS

%TRANS

F7.1

Transmissivity, percent transmittance

XMISS_FLAG_W

 

I1

XMISS quality flag

HAARDT

VOLTS

F8.3

CDOM Fluorometer, voltage

HAARDT_FLAG_W

 

I1

HAARDT quality flag

FLUORO

VOLTS

F8.3

Fluorometer, voltage

FLUORO_FLAG_W

 

I1

Fluorometer quality flag

PAR

VOLTS

F8.3

PAR, voltage

PAR_FLAG_W

 

I1

PAR quality flag

SPAR

VOLTS

F8.3

Surface PAR, voltage

SPAR_FLAG_W

 

I1

Surface PAR quality flag

CTD_DEP

METERS

F5.0

Depth

 

 

 

 

Quality Flags

CTD data quality flags were assigned to the CTDTMP (CTD temperature), CTDSAL (CTD salinity) and XMISS (Transmissivity) parameters as follows:

2                    Acceptable measurement.

3                    Questionable measurement. The data did not fit the station profile or adjacent station comparisons (or possibly bottle data comparisons). The data could be acceptable, but are open to interpretation.

4                    Bad measurement. The CTD data were determined to be unusable.

5                    Not reported. The CTD data could not be reported, typically when CTD salinity is flagged 3 or 4.

9                    Not sampled. No operational sensor was present on this cast

WHP CTD data quality flags were assigned to the CTDOXY (CTD O2), FLUORO (Fluorometer), PAR (PAR), SPAR (Surface PAR), and HAARDT (Haardt Fluorometer CDOM) parameter as follows:

1                    Not calibrated. Data are uncalibrated.

9                    Not sampled. No operational sensor was present on this cast. Either the sensor cover was left on or the depth rating necessitated removal.

Bottle Data

Description of 31H1HLY0402_hy1.csv file layout.

1st line

File type, here BOTTLE, followed by a comma and a DATE_TIME stamp 

YYYYMMDDdivINSwho

 

YYYY    4 digit year
MM      2 digit month
DD      2 digit day
div     division of Institution
INS     Institution name
who     initials of responsible person
 

#lines

A file may include 0-N optional lines, typically at the start of a data file, but after the file type line, each beginning with a "#" character and each ending with carriage return or end-of-line. Information relevant to file change/update history of the file itself may be included here, for example.

2nd line

Column headings.

3rd line

Units.

data lines

As many data lines may be included in a single file as is convenient for the user, with the proviso that the number and order of parameters, parameter order, headings, units, and commas remain absolutely consistent throughout a single file.

END_DATA  

The line after the last data line must read END_DATA.

other lines

Users may include any information they wish in 0-N optional lines at the end of a data file, after the END_DATA line.

Header columns

Parameter

Format  

Description notes

EXPOCODE

A12

The expedition code, assigned by the user.

SECT

A7

The SBI station specification. Optional.

STNNBR

A6

The originator's station number.

CASTNO

I3

The originator's cast number.

BTLNBR

A7

The bottle identification number.

BTLNBR_FLAG_W

I1

BTLNBR quality flag.

SAMPNO

A7

Cast number *100+BTLNBR. Optional

DATE

I8

Cast date in YYYYMMDD integer format.

JULIAN

F8.4

Julian day and time as fraction of day of the bottle trip.

LATITUDE

F8.4

Latitude as SDD.dddd where "S" is sign (blank or missing is positive), DD are degrees, and dddd are decimal degrees. Sign is positive in northern hemisphere, negative in southern hemisphere

LONGITUDE   

F9.4

Longitude as SDDD.dddd where "S" is sign (blank or missing is positive), DDD are degrees, and dddd are decimal degrees. Sign is positive for "east" longitude, negative for "west" longitude

DEPTH

I5

Reported depth to bottom. Preferred units are "meters" and should be specified in Line 2. In general, corrected depths are preferred to uncorrected depths. Documentation accompanying data includes notes on methodology of correction. Optional.

 

 

 

 

Parameter names, units, and comments:

Parameter 

Units  

Format  

Comments

CTDPRS

DB

F9.1

CTD pressure, decibars

CTDTMP

ITS-90

F9.3

CTD temperature, degrees C, (ITS-90)

CTDTMP_FLAG_W

 

I1

CTDTMP quality flag

CTDCOND

MS/CM

F9.3

CTD Conductivity, milliSiemens/centimeter

CTDCOND_FLAG_W

 

I1

CTDCOND quality flag

CTDSAL

 

F9.3

CTD salinity

CTDSAL_FLAG_W

 

I1

CTDSAL quality flag

SALNTY

 

F9.4

bottle salinity

SALNTY_FLAG_W

 

I1

SALNTY quality flag

CTDOXY

UMOL/KG

F9.1

CTD oxygen, micromoles/kilogram

CTDOXY_FLAG_W

 

I1

CTDOXY quality flag

CTDOXY

ML/L

F9.3

CTD oxygen, milliliters/liter

CTDOXY_FLAG_W

 

I1

CTDOXY quality flag

OXYGEN

UMOL/KG

F9.1

bottle oxygen

OXYGEN_FLAG_W

 

I1

OXYGEN quality flag

OXYGEN

ML/L

F9.3

bottle oxygen, milliliters/liter

OXYGEN_FLAG_W

 

I1

OXYGEN quality flag

O2TEMP

DEGC

F9.1

Temperature of water from spigot during oxygen draw, degrees C

SILCAT

UMOL/KG

F9.2

SILICATE, micromoles/kilogram

SILCAT_FLAG_W

 

I1

SILCAT quality flag

SILCAT

UMOL/L

F9.2

SILCATE, micromoles/liter

SILCAT_FLAG_W

 

I1

SILCAT quality flag

NITRAT

UMOL/KG

F9.2

NITRATE, micromoles/kilogram

NITRAT_FLAG_W

 

I1

NITRAT quality flag

NITRAT

UMOL/L

F9.2

NITRATE, micromoles/liter

NITRAT_FLAG_W

 

I1

NITRAT quality flag

NITRIT

UMOL/KG

F9.2

NITRITE, micromoles/kilogram

NITRIT_FLAG_W

 

I1

NITRIT quality flag

NITRIT

UMOL/L

F9.2

NITRITE, micromoles/liter

NITRIT_FLAG_W

 

I1

NITRIT quality flag

PHSPHT

UMOL/KG

F9.2

PHOSPHATE, micromoles/kilogram

PHSPHT_FLAG_W

 

I1

PHSPHT quality flag

PHSPHT

UMOL/L

F9.2

PHOSPHATE, micromoles/liter

PHSPHT_FLAG_W

 

I1

PHSPHT quality flag

NH4

UMOL/KG

F9.2

AMMONIUM, micromoles/kilogram

NH4_FLAG_W

 

I1

NH4 quality flag

NH4

UMOL/L

F9.2

AMMONIUM, micromoles/liter

NH4_FLAG_W

 

I1

NH4 quality flag

UREA

UMOL/KG

F9.2

UREA, micromoles/kilogram

UREA_FLAG_W

 

I1

UREA quality flag

UREA

UMOL/L

F9.2

UREA, micromoles/liter

UREA_FLAG_W

 

I1

UREA quality flag

FLUORO

VOLTS

F8.3

Fluorometer, voltage

FLUORO_FLAG_W

 

I1

Fluorometer quality flag

PAR

VOLTS

F8.3

PAR, voltage

PAR_FLAG_W

 

I1

PAR quality flag

SPAR

VOLTS

F8.3

Surface PAR, voltage

SPAR_FLAG_W

 

I1

Surface PAR quality flag

N**

UMOL/L

F9.2

N**, micromoles/liter

N**_FLAG_W

 

I1

N** quality flag

CHLORO

UG/L

F7.2

Chlorophyll, micrograms/liter

CHLORO_FLAG_W

 

I1

Chlorophyll quality flag

PHAEO

UG/L

F7.2

Phaeophytin, micrograms/liter

PHAEO_FLAG_W

 

I1

Phaeophytin quality flag

BTL_DEP

METERS

F5.0

bottle depth, meters

BTL_LAT

 

F8.4

Latitude at time of bottle trip, decimal degrees

BTL_LONG

 

F9.4

Longitude at time of bottle trip, decimal degrees

 

 

 

 

Quality Flags

CTD data quality flags were assigned to CTDPRS (CTD pressure), CTDTMP (CTD temperature), CTDCOND (CTD Conductivity), and CTDSAL (CTD salinity) as defined in Data Distribution, CTD Data, Quality Flags section of this document. CTDOXY (CTD O2), FLUORO (Fluorometer), PAR (PAR), and SPAR (Surface PAR) parameters are flagged with either a 2, acceptable or 9, not drawn.

Bottle quality flags were assigned to the BTLNBR (bottle number) as defined in the WOCE Operations Manual [Joyce] with the following additional interpretations:

2                    No problems noted.

3                    Leaking.  An air leak large enough to produce an observable effect on a sample is identified by a flag of 3 on the bottle and a flag of 4 on the oxygen.  (Small air leaks may have no observable effect, or may only affect gas samples.)

4                    Did not trip correctly.  Bottles tripped at other than the intended depth were assigned a flag of 4.  There may be no problems with the associated water sample data.

9                    The samples were not drawn from this bottle. 

WHP water sample quality flags were assigned to the water samples using the following criteria:

1                    The sample for this measurement was drawn from the water bottle, but the results of the analysis were not (yet) received.

2                    Acceptable measurement.

3                    Questionable measurement. The data did not fit the station profile or adjacent station comparisons (or possibly CTD data comparisons). No notes from the analyst indicated a problem. The data could be acceptable, but are open to interpretation.

4                    Bad measurement. The data did not fit the station profile, adjacent stations or CTD data. There were analytical notes indicating a problem, but data values were reported. Sampling and analytical errors were also flagged as 4.

5                    Not reported. The sample was lost, contaminated or rendered unusable.

9                    The sample for this measurement was not drawn.

Not all of the quality flags are necessarily used on this data set.


 

 

References

Armstrong, F. A. J., Stearns, C. R., and Strickland, D. H., “The measurement of upwelling and subsequent biological processes by means of the Technicon AutoAnalyzer and associated equipment,” Deep-Sea Research, 14, pp. 381-389, (1967).

Bernhardt, Wilhelms A., “The continuous determination of low level iron, soluble phosphate and total phosphate with the AutoAnalyzer”, Technicon Symposia, I, pp. 385-389 (1967).

Carpenter, J. H., “The Chesapeake Bay Institute technique for the Winkler dissolved oxygen method,” Limnology and Oceanography, 10, pp. 141-143 (1965).

Culberson, C. H., Knapp, G., Stalcup, M., Williams, R.T., and Zemlyak, F., “A comparison of methods for the determination of dissolved oxygen in seawater,” Report WHPO 91-2, WOCE Hydrographic Programme Office (Aug 1991).

Gordon, L.I., Jennings, J.C., Ross, A.A. and J.M. Krest, “A Suggested Protocol for Continuous Flow Automated Analysis of Seawater Nutrients in the WOCE Hydrographic Program and the Joint Global Ocean Fluxes Study”. 1993. WOCE Hydrographic Programs Office, Methods Manual WHPO 91-1.

Intergovernmental Oceanographic Commission, Scientific Committee on Oceanic Research Manual and Guides 29 Protocols for the Joint Global Ocean Flux Study (JGOFS) Core Measurements.  UNESCO, 170pp., (1994).

Joyce, T. ed., and Corry, C. ed., “Requirements for WOCE Hydrographic Programme Data Reporting,” Report WHPO 90-1, WOCE Report No. 67/91 3.1, pp. 52-55, WOCE Hydrographic Programme Office, Woods Hole, MA, USA (May 1994, Rev. 2), UNPUBLISHED MANUSCRIPT

Macdonald, R.W., F. A. McLaughlin and C.S. Wong, “The storage of reactive silicate samples by freezing,” Limnology and Oceanography, 31, pp. 1139-1142 (1986).

Patton, C.J. and Crouch, S.R., “Spectrophotometric and kinetics investigation of the Berthelot reaction for the determination of ammonia,” Analytical Chemistry, 49(3), pp.464-469 (1977).

Rahmatullah, Mohammed, and Boyde, T.R.C, “Improvements in the determination of urea using diacetyl monoxime; methods with and without deproteinisation,” Clinica Chimica Acta, 107, pp.3-9 1980.

Sea-Bird Electronics, Inc, CTD Data Acquisition Software Manual, March 2001.

Sea-Bird Electronics, Inc., CTD Operating and Repair Manual, February 2002.